Post-market

Revised 2026-10-08.

This page is the post-market procedure Cadasto B.V. runs as manufacturer of every FerroHEALTH product: how to complain or report an incident, how a report is classified, what happens when a released version turns out not to conform, the reports to the authorities, and what happens if the manufacturer ceases operations. The duties come from two regulations: Regulation (EU) 2025/327 on the European Health Data Space (the EHDS), Articles 30, 43, 44 and 45, and Regulation (EU) 2024/2847, the Cyber Resilience Act (the CRA), Articles 13 and 14. The steps that carry them out are the manufacturer’s own design.

Each product keeps its own registers and the steps bound to its own repository (its tracker, its release lane, its withdrawal tooling), and its book lists what counts as a possible serious incident for that product.

Warning

The quotations are from the Official Journal texts at EUR-Lex: the EHDS, the CRA, the NIS2 Directive, Regulation (EU) 2019/1020 and Commission Delegated Regulation (EU) 2026/881. The exact texts these pages were checked against are vendored in the FerroEHR repository under docs/law/eu/. This page is not legal advice, and it says nothing about whether a product or a deployment meets either regulation. It says what the manufacturer does.

Which products the duties reach

The CRA reaches every product: each tagged release is a product with digital elements that Cadasto B.V. places on the market. The EHDS duties on this page bind the manufacturer of an EHR system. Cadasto B.V. declares two:

This book records no EHDS classification for FerroTERM or FerroCHART.

When the duties apply

DutyApplies fromText
CRA Art. 14, reporting actively exploited vulnerabilities and severe incidents11 September 2026, for every release, including those published before 11 December 2027CRA Art. 71(2), Art. 69(3)
EHDS Art. 30 (complaints, registers, corrective action), Art. 43 to 45 (market surveillance, serious incidents)26 March 2027EHDS Art. 105, which gives none of them a later date
CRA Art. 13(6), 13(21), 13(23) and Annex I Part II (upstream reporting, corrective measures, cessation, vulnerability handling)11 December 2027CRA Art. 71(2)

The procedure runs now for all of them, so that it is practised before the EHDS and the rest of the CRA apply.

Who does what

WhoWhat
Cadasto B.V., the manufacturerReceives complaints and incident reports at info@cadasto.com, the single point of contact (EHDS Art. 30(1)(g)). Decides whether an event is a serious incident or an actively exploited vulnerability, and signs every report to an authority.
The maintainer, Ruben TalstraReads private vulnerability reports, the reports Cadasto B.V. forwards, and the trackers; enters the register rows, prepares the correcting release, withdraws a release, and drafts the advisory and the reports for Cadasto B.V. to send.
The deploying organisationSupplies the facts of its deployment. Owes its own notifications under the GDPR and, where it is an essential or important entity, under NIS2. None of the manufacturer’s reports stands in for them.

Making a complaint or a report

EHDS Art. 30(1)(n) asks the manufacturer to “establish channels of complaint and keep distributors informed thereof”. Pick the channel by what the report contains, and put the product’s name in the subject:

WhatChannelWho reads it
A vulnerability, including one you have seen exploitedGitHub private vulnerability reporting on the product’s repository, or info@cadasto.com with “Product vulnerability” in the subject (security)the maintainer; an email is forwarded by Cadasto B.V. the same day
Anything that harmed a person, or could haveinfo@cadasto.com, subject “Product incident”Cadasto B.V., forwarded to the maintainer the same day
Any other complaintinfo@cadasto.com, subject “Product complaint”, or a public GitHub issue on the product’s repositoryCadasto B.V. and the maintainer

Say which product and which version you run, and how it is deployed, and describe what happened. Never send patient data: describe the case, or build a synthetic one. A product may offer a command that writes a report of the deployment for this purpose; its book says which.

An emailed vulnerability report runs through the same procedure as a private GitHub report. Its arrival is the moment the manufacturer becomes aware for CRA Art. 14, so whoever reads info@cadasto.com forwards a message about a vulnerability or an exploit to the maintainer at once and writes down the hour it arrived.

How a report is classified

ClassWhen
complaintthe product behaves as documented, and the report asks for something else
non-conformitya released version misses an EHDS Annex II item, an obligation of EHDS Chapter III or a CRA Annex I requirement; continue with corrective action
serious incidentthe event meets EHDS Art. 2(2)(r); start the serious-incident clock at once
exploited vulnerabilitya vulnerability with reliable evidence of exploitation (CRA Art. 3(42)), or a severe incident (CRA Art. 14(5)); start the CRA clock at once

The person who reported is answered through the channel they used.

The registers

EHDS Art. 30(1)(o) asks the manufacturer to “keep a register of complaints and a register of non-conforming EHR systems and keep distributors informed thereof”. CRA Art. 13(6) adds the record of vulnerabilities reported upstream. The registers are kept per product, public, in the product’s repository, appended and never rewritten:

ProductRegisters
FerroEHRcomplaints, non-conforming versions, upstream reports
FerroFEDcomplaints, non-conforming versions
FerroCHART, FerroTERM, FerroBRIDGEnone yet

A complaint gets its row within one working day of receipt, whatever the channel. No row names the person who complained or carries patient data, and the upstream register names components, never people. A vulnerability enters the registers when its advisory is published, so a row never discloses one before its fix; until then the draft advisory is the record.

Corrective action, withdrawal and recall

When the manufacturer considers, or has reason to believe, that released versions “are not or are no longer in conformity with the essential requirements laid down in Annex II”, it takes “without undue delay any necessary corrective action”, or recalls or withdraws them (EHDS Art. 30(1)(i)). CRA Art. 13(21) asks for the same “immediately” where a product or the manufacturer’s processes do not conform with the CRA’s Annex I.

A withdrawal is any measure that prevents a product in the supply chain from being made available; a recall is any measure that achieves the return of a product already made available to the end user (Regulation (EU) 2019/1020 Art. 3(22) and (23), which EHDS Art. 2(1)(d) and CRA Art. 3(49) and (50) adopt). The steps:

  1. The finding enters the product’s register of non-conforming versions.
  2. The fix ships in a new patch release. A published release is immutable, so a non-conforming version is never changed or deleted.
  3. The manufacturer decides between correction, withdrawal and recall, and writes the decision and its timetable into the register. A withdrawal moves the product’s floating image tags (<major>.<minor> and latest) to the correcting release and lists the version as unsupported in the product’s SECURITY.md. The version’s own tag and image digests stay published, so a deployment pinned to them keeps running the withdrawn version until it moves. A recall adds a direct message to every known user to replace the version. FerroEHR and FerroFED each carry scripts/release/withdraw.sh for the steps bound to their own repository.
  4. The national authorities of each Member State where the version was made available or put into service are told of the non-conformity, of the corrective action “including the timetable for implementation”, and of the date the version was “brought into conformity or been recalled or withdrawn” (EHDS Art. 30(1)(i)).
  5. Distributors, the authorised representative, importers and users are told of “the non-conformity and of any corrective action, recall or withdrawal” (EHDS Art. 30(1)(j)): through a security advisory or an announcement, the release notes of the correcting version, and directly where the manufacturer knows them.

The same routes carry any “mandatory preventive maintenance of the EHR systems and its frequency” (EHDS Art. 30(1)(k)).

Serious incidents (EHDS Art. 44)

EHDS Art. 2(2)(r) defines a serious incident as “any malfunction or deterioration in the characteristics or performance of an EHR system made available on the market that directly or indirectly leads, might have led or might lead to any of the following: (i) the death of a natural person or serious harm to a natural person’s health; (ii) serious prejudice to a natural person’s rights; (iii) serious disruption of the management and operation of critical infrastructure in the health sector”. Each EHR system’s book lists the events its manufacturer treats as possible serious incidents: FerroEHR’s and FerroFED’s.

EHDS Art. 44(7) sets the report:

  • To whom: “the market surveillance authorities of the Member States where such serious incident occurred and of the Member States where such EHR systems are placed on the market or put into service”. Each Member State designates its authority, and “The Commission and the Member States shall make that information publicly available” (Art. 43(2)).
  • What: the incident, including “a description of the corrective action taken or envisaged by the manufacturer”.
  • When: “immediately after the manufacturer has established a causal link between the EHR system and the serious incident or the reasonable likelihood of such a link and, in any event, not later than three days after the manufacturer becomes aware of the serious incident involving the EHR system”. The three days run from awareness, so the report goes in on time while the cause is still being established, and is completed later.

The steps:

  1. Day 0, on awareness. The date and hour are written down, and the complaint and non-conformity rows are entered. The deployment is asked how it is deployed and what happened.
  2. The same day. The manufacturer checks whether the event is also a CRA Art. 14 matter. If it is, that clock runs as well; neither report replaces the other.
  3. By day 3 at the latest. Cadasto B.V. reports to the authorities above: what happened, the versions involved, the causal link as far as it is established, and the corrective action taken or envisaged.
  4. Harm. Where an authority finds that an EHR system “has caused harm to the health or safety of natural persons”, the manufacturer “shall immediately provide information and documentation” to the affected person or user (Art. 44(3)).
  5. Corrective action continues as above for every copy placed on the market in the Union (Art. 44(4)).

Where a serious incident concerns personal data protection, the market surveillance authority informs the data protection supervisory authorities (Art. 44(6)); a deployment’s own duties under the GDPR remain its own.

Actively exploited vulnerabilities and severe incidents (CRA Art. 14)

The manufacturer becomes aware through a private vulnerability report, an emailed one, a CSIRT telling it of someone else’s notification (CRA Art. 15(4)), or its own finding. The deadlines and the content of each step are on the security page. The steps:

  1. Hour 0. The date and hour are written down. Cadasto B.V. decides, on the evidence, whether the vulnerability is actively exploited or the incident severe. CRA Art. 14(5) calls an incident severe when it affects or can affect the ability to protect the availability, authenticity, integrity or confidentiality of sensitive or important data or functions, or when it has led or can lead to malicious code.
  2. Within 24 hours: the early warning, through ENISA’s single reporting platform to the CSIRT designated as coordinator in the Netherlands.
  3. Within 72 hours: the notification, saying how sensitive Cadasto B.V. considers the information. Where a fix is expected within 72 hours, the notification says so: that is a condition under which the receiving CSIRT may delay passing it on (Commission Delegated Regulation (EU) 2026/881 Art. 3(a), adopted for the delay CRA Art. 16(2) allows).
  4. Users are told through a GitHub security advisory on the product’s repository and the release notes of the fixing release, with the mitigations users can apply before they upgrade (Art. 14(8)).
  5. On request, an intermediate report to the CSIRT (Art. 14(6)).
  6. The final report: for a vulnerability, no later than 14 days after a corrective or mitigating measure is available; for an incident, within one month after the notification.

An event can be an EHDS serious incident and a CRA notification at once. Each report is then made, each on its own clock.

NIS2

EHDS Art. 44(7) makes the serious-incident report “without prejudice to incident notification requirements under Directive (EU) 2022/2555”, the NIS2 Directive. NIS2 Art. 23 has each Member State require essential and important entities to notify significant incidents to their CSIRT or competent authority, and healthcare providers are a sector in its Annex I. A hospital running a FerroHEALTH product may owe that notification under its own national law. The manufacturer’s EHDS report and CRA notification do not stand in for it, and the hospital’s notification does not stand in for them. The manufacturer gives the hospital the facts it needs: the advisory, the affected versions, the mitigations and the timeline it reported.

Security advisories

CRA Annex I Part II(4) has the manufacturer, “once a security update has been made available, share and publicly disclose information about fixed vulnerabilities”. Every fixed vulnerability gets a GitHub security advisory on the product’s repository, published with the fixing release. What an advisory carries, and when its technical details may wait, is on the security page.

Vulnerabilities in integrated components

When the manufacturer identifies a vulnerability “in a component, including in an open source-component, which is integrated in the product”, it shall “report the vulnerability to the person or entity manufacturing or maintaining the component”, and share any fix it wrote (CRA Art. 13(6)).

  • Who reports: the maintainer, on behalf of Cadasto B.V.
  • When: as soon as the vulnerability is confirmed in the component, whether the product’s own work found it (a fuzz crash, a review, a test) or a reporter told the manufacturer of it. A component vulnerability that a scanner reports from a published advisory is already known upstream and needs no report.
  • Through which channel: the component’s own security policy first. For a Rust crate, its SECURITY.md or private vulnerability reporting on its repository, then a RustSec advisory once it can be public; for a base-image package, the project’s security contact or the distribution’s security tracker; for a vendored asset, its publisher’s contact.
  • The fix: a patch the manufacturer wrote is offered upstream, under the component’s licence.
  • The product’s own side: the vulnerability is handled like any other: a draft advisory, a security-only release when the product is affected, or a VEX statement when it is not.
  • The record: a row in the product’s upstream register once the upstream advisory is public, naming the component and never a person.

A request from an authority

On request, the manufacturer gives a market surveillance authority “all the information and documentation necessary to demonstrate the conformity” of the product, in an official language of the Member State concerned (EHDS Art. 30(1)(l)), “in paper or electronic form” and “in a language which can be easily understood by that market surveillance authority” (EHDS Art. 30(5); CRA Art. 13(22)), and cooperates on any action to bring the product into conformity or to eliminate its risks (EHDS Art. 30(1)(m), Art. 44(1)). A request goes to the single point of contact, info@cadasto.com. An authority may restrict, recall or withdraw an EHR system whose manufacturer does not cooperate or whose information “is incomplete or incorrect” (EHDS Art. 43(5)).

If the manufacturer ceases operations

CRA Art. 13(23): “A manufacturer that ceases its operations and, as a result, is not able to comply with this Regulation shall inform, before the cessation of operations takes effect, the relevant market surveillance authorities as well as, by any means available and to the extent possible, the users of the relevant products with digital elements placed on the market, of the impending cessation of operations.”

If that happens, Cadasto B.V. will:

  1. Decide and date it. The board of Cadasto B.V. decides that it will cease operations, or that a product will no longer be maintained, and the date it takes effect. The steps below run as soon as that date is set, so every notice goes out before it.
  2. Tell the authorities first: the market surveillance authority of each Member State where a product is made available, and, while the EHDS applies, of each Member State where an EHR system is put into service. The notice names the products, the date, the date from which vulnerabilities are no longer handled, and what stays published.
  3. Tell the known users: a direct message to every operator and every commercial licensee it knows of.
  4. Tell the public: a pinned issue on each product’s repository, a notice at the top of each product’s SECURITY.md, on its documentation site and in this book, and a final security advisory naming the last supported release. The notice says that the support periods end early, on the cessation date.
  5. Leave everything published. Every release, image, chart, crate, advisory, register and page stays where it is. Releases are immutable, and nothing is withdrawn because the manufacturer stops.
  6. Close the registers: every open row gets its outcome as it stands.

Under the licence published with each version, a version becomes licensed under the Apache License 2.0 four years after its publication (licensing). The licence text ties that change to the version’s publication date and sets no condition on the Licensor continuing to exist. A commercial licence is governed by its own contract.

Open points

Nothing in the products’ repositories answers these. Cadasto B.V. supplies them:

  1. The CSIRT designated as coordinator for the Netherlands under CRA Art. 14(7), and Cadasto B.V.’s account on ENISA’s single reporting platform (Art. 16(1)).
  2. Who at Cadasto B.V. decides that a vulnerability is actively exploited or an incident severe, and signs the CRA notification and the EHDS serious-incident report; and who stands in when that person is away, since the clock runs in hours.
  3. The market surveillance authority of each Member State where a product is placed on the market or put into service, with its contact for a serious-incident report.
  4. Whether Cadasto B.V. is itself an essential or important entity under NIS2 in the Netherlands, in particular once it hosts a product as a service for other organisations.
  5. The list of distributors, importers and known users to whom the EHDS Art. 30(1)(j) and (n) notices go, which is the register of economic operators of EHDS Art. 35.
  6. Whether info@cadasto.com should publish an encryption key for vulnerability reports, and who reads that inbox and forwards a report the same day, including outside working hours.
  7. The 30-day ceiling on delaying an advisory’s technical details, which the security page states, needs Cadasto B.V.’s confirmation.
  8. Who on the board of Cadasto B.V. decides a cessation of operations, and whether a successor would take over the support periods.
  9. Whether the certified scope of Cadasto B.V.’s ISO 9001, ISO/IEC 27001 and NEN 7510 certificates covers the development and release of the products, these procedures, and a service Cadasto B.V. hosts for other organisations.